Equipment and Calibration History
View Equipment |
Date: |
2015-02-02 |
From: |
Factory - |
To: |
Lab - none |
Title: |
Returned |
Notes: |
Returned from Licor on RMA 00186861 for internal corrosion due to a leak. Licor says: Replace corroded connector LED board. We couldn't see any way that the water got inside. Very strange. We did add some water resistant grease under the spinning mirror that is now a part of how we build all instruments. The analog board had some component that was shorting out all the 5 volts for the whole instrument causing the lights to blink and motor spin. replaced the board. calibrated instrument. |
Entered by: |
Joe Verfaillie |
|
|