Date: |
2015-01-06 |
From: |
Lab - none |
To: |
Lab - none |
Title: |
Calibration |
Notes: |
Last calibration: 2014-10-02
AGC=53 SDM=7 DAC1=0-1000 DAC2=10-25
Ambient P=101.62kPa, Tair=21.21C, Tdew=5.11C
Zero Air at 950psi, God CO2 421.22 at 1300 Tdewspan=15.95
CO2 zero 0.95 -> 0.9504 Offset: 1.58ppm
CO2 span 0.9989 -> 1.0086 Offset: -5.12ppm
H2O zero 0.8505 -> 0.8510 Offset: 3.0ppm
H2O span 0.9819 -> 1.000 Offset: -0.4ppm
Notes: External serial connector on the box did not work.
|
Entered by: |
|
|
|