Equipment and Calibration History

View Equipment |

View history entry 1651 for Licor LI-7500A sn: 75H-2176

Date: 2015-01-06
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2014-09-23 AGC=90.0 SDM=1 DAC1=- DAC2=- Ambient P=101.28kPa, Tair=22.82C, Tdew=5.69C Zero Air at 950psi, God CO2 421.22 at 1300 Tdewspan=15.95 CO2 zero 0.8685 -> 0.8691 Offset: 2.56ppm CO2 span 1.0145 -> 1.0215 Offset: -3.66ppm H2O zero 0.8547 -> 0.8559 Offset: 6.6ppm H2O span 1.0075 -> 1.0099 Offset: -0.05ppm Notes:
Entered by: