View history entry 1648 for Licor LI-7500A sn: 0041

Date: 2014-12-22
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2014-06-04 AGC=51 SDM=7 DAC1=0-1000 DAC2=10-25 Ambient P=101.23kPa, Tair=22.30C, Tdew=13.20C Zero Air at 1000psi, God CO2 421.22 at 1300 Tdewspan=15.95 CO2 zero 1.0009 -> 1.0008 Offset: -0.1ppm CO2 span 1.0101 -> 1.0116 Offset: -0.82ppm H2O zero 0.8235 -> 0.8242 Offset: 4.1ppm H2O span 1.0013 -> 1.1234 Offset: -2.69ppm Notes: Dew point drifting up after calibration - might have been dew point generator.
Entered by: