Date: |
2013-11-06 |
From: |
Factory - |
To: |
Factory - |
Title: |
Repair |
Notes: |
Could not upload new OS. Lost its brains while running on the tram. Campbell: Replaced PLD U11E, which was causing intermittent communication errors/watchdogs. Replaced water damaged CS I/O connector P23B. Loaded lastest OS 23. Bench calibrated and tested unit, then tested unit in environmental chamber. Retested unit at bench. All problems resolved. |
Entered by: |
Joe Verfaillie |
|
|