Date: |
2014-10-02 |
From: |
Lab - none |
To: |
Lab - none |
Title: |
Calibration |
Notes: |
Last calibration: 2014-06-24
AGC=54 SDM=7 DAC1=0-1000 DAC2=10-25
Ambient P=100.76kPa, Tair=23.23C, Tdew=14.05C
Zero air at 1000psi, God CO2 421.22 at 1300 Tdewspan=15.81
CO2 zero 0.9506 -> 0.9500 Offset: -2.15ppm
CO2 span 1.0074 -> 0.9989 Offset: 4.78ppm
H2O zero 0.8506 -> 0.8505 Offset: -0.3ppm
H2O span 0.996 -> 0.9819 Offset: 0.3ppm
Notes: Exterior serial connector doesn't work. This calibration was done using the thermistor in the box and not on the calibration tube.
|
Entered by: |
|
|
|