Equipment and Calibration History

View Equipment |

View history entry 1566 for Licor LI-7500A sn: 75H-2176

Date: 2014-09-23
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2014-03-17 AGC=88.9 SDM=1 DAC1=- DAC2=- Ambient P=100.22kPa, Tair=26.52C, Tdew=16.19C Zero Air at 1050psi, God CO2 421.22 at 1350 Tdewspan=15.80 CO2 zero 0.8687 -> 0.8685 Offset: -0.73ppm CO2 span 1.0113 -> 1.0145 Offset: -1.74ppm H2O zero 0.8531 -> 0.8547 Offset: 8.8ppm H2O span 0.9904 -> 1.0075 Offset: -0.37ppm Notes:
Entered by: