View history entry 1560 for Licor LI-7500A sn: 0035

Date: 2014-09-09
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2014-03-17 AGC=50 SDM=7 DAC1=0-1000 DAC2=10-25 Ambient P=100.49kPa, Tair=23.33C, Tdew=13.88C Zero Air at 1000psi, God CO2 421.22 at 1300 Tdewspan=15.25 CO2 zero 0.9389 -> 0.9386 Offset: -1.05ppm CO2 span 1.0139 -> 1.0181 Offset: -3.02ppm H2O zero 0.8932 -> 0.8957 Offset: 20ppm H2O span 0.9462 -> 0.9838 Offset: -0.79ppm Notes: The signal seems a little noisy: +-1ppm CO2 and +-20mmol/m3 H2O
Entered by: