Date: 2014-09-10
From: East End - Gas Analyzer CH4
To: East End - Gas Analyzer CH4
Title: Calibration
Notes: Calibration Location: East End Last calibration: 2014-03-07 RSSI=80%, Optical RH=23% Ambient P=101.23kPa, Tair=29.9C Zero Air at 1400psi, 2nd CH4 4.995 at 700 CH4 Zero Offset: 0.085ppm CH4 Span Offset: -0.195ppm Notes: Very hot day with no wind. Offsets larger than usual.