Equipment and Calibration History

View Equipment |

View history entry 1491 for Los Gatos Research FMA RMT-200 sn: 08-0114

Date: 2011-08-12
From: Twitchell Rice -
To: Twitchell Rice -
Title: Calibration
Notes: The LGR pressure was really low 89.28 and in slow flow it was even lower. I checked the external 2um filter and it seemed clean. I checked the internal 2um filter and it was completely stopped up. I replaced it with a cleaned one. I blew the replacement filter clear with the cal tank but it still brought the ring time down after I got it all back together. The pressure was better and I had to close down the bypass valve to bring it in range. 9:00PST, 24.16C, 144.56Torr, 13.300usec, 10Hz, 1.831ppm The low flow pressure was still a bit low during calibration: 9:06PST, 24.05C, 125.31Torr, 13.302usec, 1Hz, 3.270ppm The calibration was almost right on with the 3.267ppm tank at 650psi. Last cal was July 1, 2011. The pressure fell to 111.7 by the end of the calibration. I also replace the 40um filter strainer. It was very dirty and had some spider parts in it. Before leaving the LGR read: 9:34PST, 24.55C, 144.81Torr, 13.291usec, 10Hz, 1.827ppm
Entered by: Sara Knox
Field notes