Date: |
2010-12-02 |
From: |
Twitchell Rice - |
To: |
Twitchell Rice - |
Title: |
Calibration |
Notes: |
Pre cal:
7:43PST, 24.90C, 142.80Torr, 16.064usec, 1Hz, 2.660ppm - off by 616ppb a lot.
Post cal CH4 was 3.263ppm
And back in operation after adjusting the bypass valve because pressure was too high:
7:53PST, 23.33C, 143.96Torr, 16.089usec, 10Hz, 3.266ppm |
Entered by: |
Sara Knox |
|
|