Date: |
2010-09-08 |
From: |
Twitchell Rice - |
To: |
Twitchell Rice - |
Title: |
Calibration |
Notes: |
I adjusted the bypass valve. When Frank arrived we did a calibration. Last cal was done on Aug 6, 2010. The field tank is 3.276ppm CH4
Precal: 9:52PST, 25.54C, 111.2Torr, 16.112usec, 1Hz, 3.377ppm
Low flow pressure seems low here too but I did not check the filters. Being 100ppb off seems a lot, last cal uses a standard closer to 1.8ppm though. Post cal CH4 read 3.276ppm.
When we left the LGR read:
10:02PST, 25.49C, 144.34Torr, 16.080usec, 10Hz, 1.840ppm |