View history entry 1483 for Los Gatos Research FGGA sn: 09-0010

Date: 2010-05-20
From: Twitchell Rice -
To: Twitchell Rice -
Title: Calibration
Notes: calibrated the LGR: Initial readings: 9:35, 19.99C, 138.24Torr, 14.071usec, 10Hz, 1.8530ppm, 8187ppm, 406.97ppm bypass valve adjusted to 143.55Torr. Calibrated with 3.276ppm CH4, 433.0ppm CO2 tank with 1800psi Pre cal: 9:41, 138.16Torr, 1Hz, 3.220+-0.005ppm, 444+-1ppm Post cal: 9:48, 3.278+-0.002ppm, 432+-1ppm Ambient conditions after calibration: 9:50, 20.02C, 143.90Torr, 14.069usec, 10Hz, 1.8610ppm, 8332ppm, 390.55ppm
Entered by: Sara Knox