View history entry 1481 for Los Gatos Research FGGA sn: 09-0010

Date: 2010-04-22
From: Twitchell Rice -
To: Twitchell Rice -
Title: Calibration
Notes: The LGR would not line lock CH4/H2O when power was restored. Initially the cell temperature was about 10C and I let it warm up to 18C with no lock and many restarts. Finally I called LGR and got their maintenance password to change line lock settings from 0.165 to 0.185 and this did the trick. Initial readings 11:35, 18.50C, 144.60Torr, 13.940usec, 10Hz, 2.149ppm, 12039ppm, 398.07ppm Calibrated with 3.276ppm CH4, 433.0ppm CO2 tank, last cal Mar 11, 2010 Precal 11:40, 138.24Torr, 3.270+-0.002ppm CH4, 425.0+-1ppm CO2 Postcal: 11:44, 138Torr, 3.279+-0.002ppm CH4, 433.5+-1ppm CO2