Last calibration: 2009-03-03 00:00:00
Cell Pressure=148.63kPa, Cell Temperature=28.27C, Ring Time=17.277usec
2nd Standard at 0psi: CH4: 3.276ppm, CO2: 433ppm, H2O: 0ppm
CH4 = 3.368ppm -> 3.275ppm Offset: -0.092ppb
CO2 = 432.1ppm -> 433.4ppm Offset: 0.9ppm
Notes: LGR pressure very low ~16Torr, but very easily adjusted using bypass valve. |