Date: 2013-01-08
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Trueppm = Probeppm*0.9828 + 15.78; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 421ppm, 3056ppm, and 10000ppm. Ta=22.3C, P=102.5kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Sara.