Equipment and Calibration History

View Equipment |

View history entry 1402 for Vaisala GMP221 sn: Y0710013

Date: 2012-02-03
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Trueppm = Probeppm*1.07 + -227; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 436ppm, 2238ppm, and 10000ppm (N2 bal). Ta=26.6C, P=101.6kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Siyan.
Entered by: Joe Verfaillie