Date: |
2011-12-13 |
From: |
Lab - none |
To: |
Lab - none |
Title: |
Calibration |
Notes: |
Trueppm = Probeppm*1.039 + 10.06; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 406ppm, 2238ppm, and 10000ppm (N2 bal). Ta=24.6C, P=101.7kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Sara. |
Entered by: |
Joe Verfaillie |
|
|