Equipment and Calibration History

View Equipment |

View history entry 1392 for Vaisala GMT221 sn: F4960019

Date: 2011-10-06
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Trueppm = Probeppm*1.02 + -56.65; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 406ppm, 2238ppm, and 10000ppm (N2 bal). Ta=26.2C, P=101.4kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Sara.
Entered by: Joe Verfaillie