Date: |
2011-03-14 |
From: |
Lab - none |
To: |
Lab - none |
Title: |
Calibration |
Notes: |
Trueppm = Probeppm*1.046 + 56.59; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 597ppm, 2238ppm, and 10000ppm (N2 bal). Ta=25.3C, P=102.5kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Melanie. |
Entered by: |
Joe Verfaillie |
|
|