Equipment and Calibration History

View Equipment |

View history entry 1377 for Vaisala GMT222 sn: G0230062

Date: 2011-02-07
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Trueppm = Probeppm*1.075 + 6.9; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 597ppm, 2238ppm, and 10000ppm (N2 bal). Ta=25.8C, P=102.5kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Melanie.
Entered by: Joe Verfaillie