Date: |
2014-03-07 |
From: |
East End - Gas Analyzer CH4 |
To: |
East End - Gas Analyzer CH4 |
Title: |
Calibration |
Notes: |
Calibration Location: East End
Last calibration: 2013-03-20
RSSI=85%, Optical RH=19%
Ambient P=102.8kPa, Tair=16.3C
Zero Air at 1600psi, 2nd CH4 4.995 at 1100
CH4 Zero Offset: 0.03ppm
CH4 Span Offset: 0.015ppm
Notes: Calibrated by Cove and Sara. Firmware upgraded to v1.0.21 before cal. Windy day. A little fluctuation in span 5.01-4.995
|
Entered by: |
|
|
|