Date: |
2014-03-07 |
From: |
West Pond - Gas Analyzer CH4 |
To: |
West Pond - Gas Analyzer CH4 |
Title: |
Calibration |
Notes: |
Calibration Location: West Pond
Last calibration: 2013-08-21
RSSI=73.8%, Optical RH=7%
Ambient P=102.4kPa, Tair=20.7C
Zero Air at 1600psi, 2nd CH4 4.995 at 1100
CH4 Zero Offset: 0.01ppm
CH4 Span Offset: 0.001ppm
Notes: Calibrated by Cove and Sara after firmware upgrade to v 1.0.21. Windy day. Windy day.
|
Entered by: |
|
|
|