Equipment and Calibration History

View Equipment |

View history entry 113 for Licor LI-7500A sn: 0041

Date: 2010-05-27
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last cal 2010-02-09, SMD 7 DAC1 H2O 0-1000 DAC2 CO2 10-25 AGC = 54%, T = 18.76C, P = 100.02kPa, Td = 15.4C N2 = 0.0 500psi CO2 = 597.4ppm <300psi CO2 zero = 0.9881-> 0.9882, offset +0.3ppm CO2 span = 1.0022-> 0.9984, offset +3.1ppm H2O zero = 0.7395-> 0.7385, offset -0.095mmol/m3 H2O span = 0.9928-> 0.9804, offset +0.35C
Entered by: