Equipment and Calibration History

View Equipment |

View history entry 1095 for Licor LI-7500A sn: 0035

Date: 2013-11-06
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2013-04-23 AGC=49 SDM=7 DAC1=0-1000 DAC2=10-25 Ambient P=101.35kPa, Tair=22.84C, Tdew=4.69C Zero Air at 1500psi, God CO2 421.22 at 1600 Tdewspan=15.16 CO2 zero 0.9386 -> 0.9393 Offset: 2.95ppm CO2 span 1.0156 -> 1.0154 Offset: 0.03ppm H2O zero 0.8981 -> 0.8904 Offset: -41.3ppm H2O span 0.9842 -> 0.9744 Offset: 0.17ppm Notes:
Entered by: