Equipment and Calibration History

History for Pro-Oceanus MiniCH4 38-572-25

Rec No.DateLoc FromTitleLoc ToNotes
7161 2025-03-13 Lab none Tested Lab none I ran this sensor in continuous mode for 10s of minutes and it never stopped reading -10 for CH4. I removed the filter and didn't seen any obvious signs in water getting in. Need to contact ProOcenanus.
7157 2025-03-07 Mayberry CH4 Probe Removed Lab none This sensor had fallen out of the cable ties and its sensor may have been in the sediment. The CH4 readings steadily rose past 4% but it was dropping out before climbing. Removed for testing.
7152 2025-01-15 Lab none Installed Mayberry CH4 Probe Installed on the northeast scaffolding foot at a fixed height. Interface box in the data logger enclosure.