Instrument:CH4 Probe
Manufacture:Pro-Oceanus
Model No:MiniCH4
Serial No:38-573-25
Tag No:
Acquisition Date:2025-01-03
Project:Borrowed
Location on 2025-05-16:Lab,
Notes: Borrowed from Sheel Bansal at the USGS after factory re-calibration at Pro-Oceanus. 5 probes, 3 interface boxes (2 long cable, 1 short), 2 Sea-Bird pumps, 4 external battery cases, 8 spare battery packs (8 D-cell), 4 short water proof cables, 4 wall warts, 2 USB-serial adapters, 3 serial cables, bag of port plugs.

Documents

No documents found

History

Rec No.DateLoc FromTitleLoc ToNotes
7234 2025-05-16 Lab none Tested Lab none This sensor had good numbers today. Not sure what to do with it.
7199 2025-04-24 Lab none Tested Lab none I tested this sensor in the lab. The serial interface and temperature seem fine but the CH4 is reading -10.00. I downloaded all the data and erased the internal data file. I will try to get advice from Pro-Oceanus.
7198 2025-04-17 Gilbert Tract CH4 Probe Tower Removed Lab none Removed from float at GT boardwalk after animals chewed its cable, it also started to read -10.00 % CH4.
7107 2025-01-15 Lab none Installed Gilbert Tract none Installed on a floating frame with an external battery pack on top so that the CH4 probe is horizontal just below the surface and the battery pack is horizontal just above the water. It's connected to the boardwalk with a yellow plastic chain. Should be collecting an average of 5 readings every 15 minutes.