Equipment and Calibration History

View Equipment |

View history entry 1051 for Licor LI-7700 sn: TG1-0224

Date: 2013-08-21
From: Mayberry - Gas Analyzer CH4
To: Mayberry - Gas Analyzer CH4
Title: Calibration
Notes: Calibration Location: Mayberry Last calibration: 2013-02-21 RSSI=82%, Optical RH=8% Ambient P=101.0kPa, Tair=24.2C Zero Air at 1700psi, CH4 in air 4.995 at 1300 CH4 Zero Offset: 0.000ppm CH4 Span Offset: -0.026ppm Notes: It was windy but seemed to have no effect. The concentration of the secondary ch4 standard has been reevaluated since last cal. Last time 4.982, now 4.995
Entered by:
Field notes