Date: |
2013-08-21 |
From: |
West Pond - Gas Analyzer CH4 |
To: |
West Pond - Gas Analyzer CH4 |
Title: |
Calibration |
Notes: |
Calibration Location: West Pond
Last calibration: 2013-02-19
RSSI=78%, Optical RH=3%
Ambient P=101.1kPa, Tair=21.8C
Zero Air at 1700psi, CH4 in air 4.995 at 1300
CH4 Zero Offset: 0.01ppm
CH4 Span Offset: 0.013ppm
Notes:
|