Equipment and Calibration

View History

Equipment Entry

Instrument:Gas Analyzer
Manufacture:Licor
Model No:LI-7500RS
Serial No:75H-1653
Tag No:
Acquisition Date:2024-10-24
Project:USGS-RR
Location on 2025-05-07:Loaned,
Notes: Acquired with the handoff of Rush Ranch. No temperature sensor.

Documents

No documents found

History

Rec No.DateLoc FromTitleLoc ToNotes
7222 2025-05-07 Lab none Loaned Loaned none Loaned to Delfina Grinspan for use this summer in vineyards in Sonoma and Napa.
7193 2025-04-21 Lab Calibration Lab Last calibration: 2023-09-08
AGC=101.3 SDM=1 DAC1=- DAC2=-
AIU sn AIU-2023, Presure A0=59.172, A1=15.673
Ambient P=100.4kPa, Tair=19.7C, Tdew=11.0C
Zero Air at 200psi, AMP FB04304 401.79 at 900 Tdewspan=16.0
CO2 zero 1.1420 -> 1.403 Offset: -3.7ppm
CO2 span 0.9691 -> 0.9969 Offset: -14.39ppm
H2O zero 1.0079 -> 1.0088 Offset: 4.1ppm
H2O span 1.0189 -> 1.0211 Offset: -0.05ppm
Notes: First cal after recieving from USGS with Rush Ranch stuff. Larger offsets not unexpected. We actually had this IRGA in 2011 from Franks as a loaner for Vaira.
6988 2024-10-24 Rush Ranch Gas Analyzer CO2/H2O Removed Lab none Removed from the main tower at Rush Ranch with the handoff of the site. To be cleaned up, tested and likely returned to the tower.