Equipment and Calibration
Equipment history at Bouldin Alfalfa| ALL | | Anemometer || Barometer || Camera || Data Logger || Flux Analyzer || Forced Diffusion || Gas Analyzer || Heat Flux Plate || Infrared Temperature || Interface || Multiplexer || Narrow Band || Net Radiometer || PAR Sensor || Precipitation || Range Finder || RH & Temp || Soil Moisture Probe || Solar Controller || Solar Panel || Weather Station |
SELECT a.recno, a.eh_date, a.eh_loc1, a.eh_loc2, IF(a.eh_loc2=:loc, a.eh_locref2, a.eh_locref1) AS ref, a.eh_locref2, a.eh_title, a.eh_notes, a.eh_id, b.serialno, b.equip, b.modelno FROM equiphist a INNER JOIN equiplist b ON a.eh_id=b.recno WHERE (a.eh_loc1=:loc1 OR a.eh_loc2=:loc2) AND b.equip=:eq ORDER BY b.serialno, a.eh_date
Date | Serial No. | Model | Action | Ref | Notes |
---|---|---|---|---|---|
RH & Temp | |||||
2020-11-19 | G4540021 | HMP60 | Installed | RH & Temp | Installed at Bouldin Alfalfa to replace sn T1250012 that needs to be returned to Siyan. |
2021-03-16 | G4540021 | HMP60 | Removed | RH & Temp | Removed from BA and brought back to the lab b/c AirT has been drifting low. it was replaced by HMP155 sn S3420870. |
2016-08-12 | L1740398 | HMP60 | Installed | RH & Temp | Installed with the initial set up of the site. |
2018-08-23 | L1740398 | HMP60 | Removed | RH & Temp | Removed from BA due to low temperature readings. Several other HMP60s have similarly failed. |
2019-10-02 | L1740398 | HMP60 | Removed | 0 | Removed from BA CH4 release experiment and brought back to the lab. |
2021-03-16 | S3420870 | HMP155 | Installed | RH & Temp | Installed at BA to replace sn G4540021, which was brought back to the lab for testing. |
2021-03-30 | S3420870 | HMP155 | Removed | RH & Temp | Removed from BA and brought back to the lab for repair. This sensor had fallen to the ground. The cable head needs rewiring. |
2021-04-15 | S3420870 | HMP155 | Installed | RH & Temp | Installed on the BA tower after reconnecting sensor to cable screw terminals in the lab. The collar has a metal set screw. |
2018-08-23 | T1250012 | HMP45AC | Installed | RH & Temp | Installed on BA to replace an HMP60 with low temperature readings. This sensor was in the TA inter-comparison T slope was 1.0462 and RH slope was 1.0369. |
2020-11-19 | T1250012 | HMP45AC | Removed | RH & Temp | Removed from Bouldin Alfalfa to be returned to Siyan. Replaced by sn G4540021 |
2021-08-04 | T1250012 | HMP45AC | Installed | none | Installed on the BA bouncer tower next to the upper sonic for the advection experiment. It was wired to U9/U10. |
2021-09-03 | T1250012 | HMP45AC | Removed | Removed from BA bouncer tower, where it was the upper T/RH sensor. It was replaced by sn T2820180. | |
2021-11-18 | T1250012 | HMP45AC | Installed | none | Installed on BA tower for an HMP45 intercomparison with the tower HMP155. |
2021-12-14 | T1250012 | HMP45AC | Removed | Removed from BA at the conclusion of the HMP intercomparison | |
2021-09-03 | T2820180 | HMP155 | Installed | none | Installed on the BA bouncer tower for the advection experiment in the upper T/RH location. |
2021-09-03 | T2820183 | HMP155 | Installed | none | Installed on the BA bouncer tower for the advection experiment in the lower T/RH location |
2021-11-05 | T2820183 | HMP155 | Moved | none | Removed from bouncer tower (lower position) at the conclusion of this year's advection experiment and installed on the main tower for an HMP intercomparison |
2021-11-18 | T2820183 | HMP155 | Removed | Removed from BA tower at the conclusion of the HMP155 intercomparison. | |
2021-09-03 | T2930114 | HMP155 | Installed | none | Installed on the main BA tower (lower location) for the advection experiment |
2021-11-05 | T2930114 | HMP155 | Moved | none | Removed from main tower (lower position) at the conclusion of this year's advection experiment and installed on the main tower for an HMP intercomparison |
2021-11-18 | T2930114 | HMP155 | Removed | Removed from BA tower at the conclusion of the HMP155 intercomparison. | |
2021-08-04 | U3030041 | HMP45AC | Installed | none | Installed on the BA bouncer tower next to the lower sonic for the advection experiment. It was wired to U11/U12. |
2021-09-03 | U3030041 | HMP45AC | Removed | Removed from BA bouncer tower, where it was the lower T/RH sensor. It was replaced by sn T2820183. | |
2021-11-18 | U3030041 | HMP45AC | Installed | none | Installed on BA tower for an HMP45 intercomparison with the tower HMP155. |
2021-12-01 | U3030041 | HMP45AC | Removed | none | Removed from the BA HMP intercomparison from wiring position DE1/2 because it was giving bad data. It was brought back to the lab for testing. |