| ALL | | Anemometer || Battery Bank || Camera || CO2 Probe || Data Logger || Forced Diffusion || Gas Analyzer || Heat Flux Plate || Interface || LED sensor || Moisture Point || Multiplexer || Narrow Band || Net Radiometer || PAR Sensor || Precipitation || Pyranometer || Radio || RH & Temp || Soil Moisture Probe || Solar Controller || Solar Panel || Weather Station |

SELECT a.recno, a.eh_date, a.eh_loc1, a.eh_loc2, IF(a.eh_loc2=:loc, a.eh_locref2, a.eh_locref1) AS ref, a.eh_locref2, a.eh_title, a.eh_notes, a.eh_id, b.serialno, b.equip, b.modelno FROM equiphist a INNER JOIN equiplist b ON a.eh_id=b.recno WHERE (a.eh_loc1=:loc1 OR a.eh_loc2=:loc2) AND b.equip=:eq ORDER BY b.serialno, a.eh_date

Date Serial No. Model Action Ref Notes
Precipitation
2010-09-02 24875-999 TR-525I-R1 Removed Precipitation This tipping bucket was removed from the Vaira tower to troubleshoot a problem that was causing the total number of tips not to be saved in the final storage location.
2010-09-10 24875-999 TR-525I-R1 Installed Precipitation Reinstalled on the Vaira tower after testing in the lab. The testing revealed that the problem was a bug in the data logger program that was recording instantaneous values instead of 30min averages.
2018-10-17 24875-999 TR-525I-R1 Calibration Precipitation Got 78 of 100 tips - adjusted screws 3 turns CW
2019-10-24 24875-999 TR-525I-R1 Calibration Precipitation I connected the tipping bucket to a CR23x and dripped 475ml of water through the bucket which should result in 100tips. We got 103 tips. I did not adjust it.
2021-10-06 24875-999 TR-525I-R1 Calibration Precipitation Tested using our new method with the peristaltic pump and 237ml of water which should result in 50 tips. I got 54 tips. I adjusted the screws two full turns CCW which should decrease the number of tips by 8-12% (4-6 tips).