Equipment history at Hill Slough| ALL | | Anemometer || Battery Bank || Camera || CO2 Probe || Conductivity Sensor || Data Logger || DO Sensor || Fluorometer || Forced Diffusion || Gas Analyzer || Heat Flux Plate || Interface || Multiplexer || Narrow Band || Net Radiometer || PAR Sensor || pH ORP || Precipitation || Range Finder || RH & Temp || Soil Moisture Probe || Solar Controller || Solar Panel || Water Level || Wiper |
SELECT a.recno, a.eh_date, a.eh_loc1, a.eh_loc2, IF(a.eh_loc2=:loc, a.eh_locref2, a.eh_locref1) AS ref, a.eh_locref2, a.eh_title, a.eh_notes, a.eh_id, b.serialno, b.equip, b.modelno FROM equiphist a INNER JOIN equiplist b ON a.eh_id=b.recno WHERE (a.eh_loc1=:loc1 OR a.eh_loc2=:loc2) AND b.equip=:eq ORDER BY b.serialno, a.eh_date
Date | Serial No. | Model | Action | Ref | Notes |
---|---|---|---|---|---|
Precipitation | |||||
2021-03-16 | 48130-611 | TE525 | Installed | Precipitation | Installed on the SE corner of the scaffolding |
2021-10-26 | 48130-611 | TE525 | Tested | Precipitation | Got 53/50 tips. I meant to turn the screws 1 turn CCW but I think I may have turned them 2 turns CCW. Needs a retest next time anyways. |
2021-11-04 | 48130-611 | TE525 | Adjusted | Precipitation | The screws were adjusted 1/2 turn in (CW) before testing. The test got 51/50 tips - okay. |