Equipment and Calibration

Equipment history at West Pond

| ALL | | Anemometer || Barometer || Battery Bank || Camera || CO2 Probe || Conductivity Sensor || Data Logger || DO Sensor || Gas Analyzer || Interface || Multiplexer || Narrow Band || Net Radiometer || PAR Sensor || RH & Temp || Solar Controller || Solar Panel || Water Level || Weather Station |

SELECT a.recno, a.eh_date, a.eh_loc1, a.eh_loc2, IF(a.eh_loc2=:loc, a.eh_locref2, a.eh_locref1) AS ref, a.eh_locref2, a.eh_title, a.eh_notes, a.eh_id, b.serialno, b.equip, b.modelno FROM equiphist a INNER JOIN equiplist b ON a.eh_id=b.recno WHERE (a.eh_loc1=:loc1 OR a.eh_loc2=:loc2) AND b.equip=:eq ORDER BY b.serialno, a.eh_date

Date Serial No. Model Action Ref Notes
Interface
2012-04-17 3586 A547 Installed Interface A547 Installed in box at west pilot pond. Not running because of forgotten power connector.
2024-02-12 3586 A547 Removed Interface A547 Removed from WP with conductivity sensor sn 6096 and brought back to the lab. Nothing wrong with the data (probably), I just wanted an extra sensor in the lab to play around with TempCoef.
2012-10-18 6761 NL115 Installed Interface NL115 ATI Installed at West Pond with the ATI sonic for a three week inter-comparison
2012-11-08 6761 NL115 Removed Interface NL115 ATI Removed from West Pond with the ATI sonic after a three week inter-comparison
2012-11-08 AIU-0397 LI-7550 Installed Interface LI-7550 Installed on the West Pond tower to swap out AIU-650 for repair of the condensation problem.
2013-01-03 AIU-0397 LI-7550 Removed Interface LI-7550 Removed from the West Pond tower for repair of the condesation problem. Replaced by AIU-0666. Heads swapped for calibration at the same time.
2013-08-21 AIU-0614 LI-7550 Installed Interface LI-7550 Installed on the West Pond tower to replace sn 0666 for clean up. This box was placed on the top level of the scaffolding to hopefully avoid the ants.
2013-11-21 AIU-0614 LI-7550 Upgrade Interface LI-7550 The firmware of the 7550 box was upgraded from v6.5.2 to v6.5.5 to fix bugs with the USB memory overwrite and synchronizing the 7700 clock.
2014-09-04 AIU-0614 LI-7550 Removed Interface LI-7550 Removed from the West Pond tower. In trying to swap the head, this box went into a reboot loop and we could not get it to complete a boot. This box also needs its connector circuit board upgraded to prevent condensation problems.
2012-05-02 AIU-0650 LI-7550 Installed Interface LI-7550 Installed on the West Pond tower at mid-level on the scaffolding.
2012-10-08 AIU-0650 LI-7550 Removed Interface LI-7550 Removed from West Pond to be returned to Licor for repair of condensation probelm. This unit had not failed yet. Replaced by sn 0397.
2013-01-03 AIU-0666 LI-7550 Installed Interface LI-7550 Installed on the West Pond tower with head 0065 to replace AIU-0397 for repair of condensation problem.
2013-08-21 AIU-0666 LI-7550 Removed Interface LI-7550 Removed from the West Pond tower. This box has been suffer an ant infestation for many weeks. It was removed to be de-ant-ified, cleaned up and tested in the lab.
2014-09-08 AIU-0666 LI-7550 Installed Interface LI-7550 Installed on the West Pond tower with head sn 75H-2180. There was now 7550-7500A pair on the tower because of issues with swapping the head on 9/4. This box has a new door gasket.
2014-10-02 AIU-0666 LI-7550 Upgrade Interface LI-7550 Updated the 7550 firmware from v7.0.1 to v7.3.0. Fixed bug that can cause the instrument to restart unexpectedly, added 7700 data to interface can view/log power Vin, other minor bug fixes.
2016-04-13 AIU-0666 LI-7550 Upgrade Interface LI-7550 Updated the 7550 firmware from v7.3.0 to v8.0.0 skiping version 7.5.0 Relevant 7.5.0 changes: Main Window is now resizable, Added serial number entry for sonics, Help update, Fixed timeout when saving config files Relevant 8.0.0 changes: Separate bandwidth settings for analog inputs and gas channels, Detection of new RS instruments, Additional calibration coefficients, Added sonic software version entry, Added chopper temp logging
2018-01-16 AIU-0666 LI-7550 Upgrade Interface LI-7550 Downgraded embedded Licor software from v8.7.5 to 8.5.0 because 8.7.5 caused mismatch problems between data columns and headers
2018-01-31 AIU-0666 LI-7550 Removed Interface LI-7550 Brought back to the lab for some TLC. It was replaced by AIU-1570.
2018-01-31 AIU-1570 LI-7550 Installed Interface LI-7550 Installed at WP to replace AIU-0666 that was removed to fix door gasket, pressure port, circuit board cover.
2021-01-17 AIU-1570 LI-7550 Upgrade Interface LI-7550 Upgraded embedded software from v8.5.0 to v8.8.32 in hopes this will solve the persistent 7700 sync issue
2024-02-28 AIU-1570 LI-7550 Removed Interface LI-7550 Removed from WP as part of its decomissioning
Export to BADM