SELECT a.recno, a.eh_date, a.eh_loc1, a.eh_loc2, IF(a.eh_loc2=:loc, a.eh_locref2, a.eh_locref1) AS ref, a.eh_locref2, a.eh_title, a.eh_notes, a.eh_id, b.serialno, b.equip, b.modelno FROM equiphist a INNER JOIN equiplist b ON a.eh_id=b.recno WHERE (a.eh_loc1=:loc1 OR a.eh_loc2=:loc2) AND b.equip=:eq ORDER BY b.serialno, a.eh_date

Date Serial No. Model Action Ref Notes
Gas Analyzer
2018-07-11 0041 LI-7500A Installed Gas Analyzer CO2/H2O Installed at Sherman Barn on the scaffolding frame in the southwest corner of the cow fence. This is the initial set up of the site.
2019-01-02 0041 LI-7500A Calibration Lab Last calibration: 2018-07-03 AGC=98 SDM=0 DAC1=0 DAC2=0 Ambient P=101.2kPa, Tair=20.1C, Tdew=3.7C Zero Air at 1400psi, AMP Air 408.97 at 875 Tdewspan=16.00 CO2 zero 0.9966 -> .9953 Offset: -4.19ppm CO2 span 1.0173 -> 1.0194 Offset: -1.25ppm H2O zero 0.8192 -> .8248 Offset: 34.0ppm H2O span 1.0285 -> 1.0312 Offset: -.07ppm Notes: H2O zero offset larger than I expected, but H2O span seems fine.
2018-10-18 0041 LI-7500A Removed Gas Analyzer CO2/H2O Removed from Sherman Barn for calibration. Replaced by sn 75H-2669.
2019-01-02 0041 LI-7500A Calibration Lab Last calibration: 2018-07-03 AGC=98 SDM=0 DAC1=0 DAC2=0 Ambient P=101.2kPa, Tair=20.1C, Tdew=3.7C Zero Air at 1400psi, AMP Air 408.97 at 875 Tdewspan=16.00 CO2 zero 0.9966 -> .9953 Offset: -4.19ppm CO2 span 1.0173 -> 1.0194 Offset: -1.25ppm H2O zero 0.8192 -> .8248 Offset: 34.0ppm H2O span 1.0285 -> 1.0312 Offset: -.07ppm Notes: H2O zero offset larger than I expected, but H2O span seems fine.
2019-01-31 75H-2180 LI-7500A Installed Gas Analyzer CO2/H2O Installed at SB to replace sn 75H-2669, which was brought back to the lab for calibration.
2019-04-23 75H-2180 LI-7500A Calibration Lab Last calibration: 2019-01-25 AGC=102 SDM= DAC1= DAC2= Ambient P=100.6kPa, Tair=24.7C, Tdew=10.9C Zero Air at 1100psi, AMP Air 408.97 at 700 Tdewspan=16.00 CO2 zero 0.9137 -> .9136 Offset: -.34ppm CO2 span 1.0262 -> 1.0266 Offset: -.2ppm H2O zero 0.8891 -> .8886 Offset: -3.0ppm H2O span 1.0199 -> 1.0236 Offset: -.07ppm Notes:
2019-04-18 75H-2180 LI-7500A Removed Gas Analyzer CO2/H2O Removed from SB and brought back to the lab for calibration. It was replaced by sn 75H-2669.
2019-04-23 75H-2180 LI-7500A Calibration Lab Last calibration: 2019-01-25 AGC=102 SDM= DAC1= DAC2= Ambient P=100.6kPa, Tair=24.7C, Tdew=10.9C Zero Air at 1100psi, AMP Air 408.97 at 700 Tdewspan=16.00 CO2 zero 0.9137 -> .9136 Offset: -.34ppm CO2 span 1.0262 -> 1.0266 Offset: -.2ppm H2O zero 0.8891 -> .8886 Offset: -3.0ppm H2O span 1.0199 -> 1.0236 Offset: -.07ppm Notes:
2019-07-25 75H-2667 LI-7500A Installed Gas Analyzer CO2/H2O Installed at SB to replace sn 75H-2669, which was brought back to the lab for calibration.
2019-11-12 75H-2667 LI-7500A Calibration Lab Last calibration: 2019-07-24 AGC=101.4 SDM= DAC1= DAC2= Ambient P=100.7kPa, Tair=26.2C, Tdew=10.9C Zero Air at 950psi, AMP CA4396 408.97 at 700 Tdewspan=16.00 CO2 zero 0.9413 -> .9419 Offset: 2.32ppm CO2 span 1.0158 -> 1.019 Offset: -1.67ppm H2O zero 0.8363 -> .8384 Offset: 12.8ppm H2O span 1.0222 -> 1.0239 Offset: -.03ppm Notes:
2019-10-31 75H-2667 LI-7500A Removed Gas Analyzer CO2/H2O Removed from Sherman Barn for calibration. Replaced by sn 75H-2668.
2019-11-12 75H-2667 LI-7500A Calibration Lab Last calibration: 2019-07-24 AGC=101.4 SDM= DAC1= DAC2= Ambient P=100.7kPa, Tair=26.2C, Tdew=10.9C Zero Air at 950psi, AMP CA4396 408.97 at 700 Tdewspan=16.00 CO2 zero 0.9413 -> .9419 Offset: 2.32ppm CO2 span 1.0158 -> 1.019 Offset: -1.67ppm H2O zero 0.8363 -> .8384 Offset: 12.8ppm H2O span 1.0222 -> 1.0239 Offset: -.03ppm Notes:
2019-10-31 75H-2668 LI-7500A Installed Gas Analyzer CO2/H2O Installed at Sherman Barn to replace sn 75H-2667 for calibration.
2020-02-11 75H-2668 LI-7500A Calibration Lab Last calibration: 2019-10-30 AGC=100.4 SDM=0 DAC1=- DAC2=- Ambient P=100.8kPa, Tair=26.9C, Tdew=-1.59C Zero Air at 800psi, AMP CA04396 408.97 at 550 Tdewspan=16.0 CO2 zero 0.9393 -> 0.9393 Offset: -0.1ppm CO2 span 1.016 -> 1.0171 Offset: -0.94ppm H2O zero 0.9096 -> 0.9072 Offset: 2.7ppm H2O span 1.0143 -> 1.0147 Offset: -0.01ppm Notes: Ambient air very dry
2020-02-06 75H-2668 LI-7500A Removed Gas Analyzer CO2/H2O Removed from Sherman Barn and brought back to the lab for calibration. Replaced by sn 75H-2669.
2020-02-11 75H-2668 LI-7500A Calibration Lab Last calibration: 2019-10-30 AGC=100.4 SDM=0 DAC1=- DAC2=- Ambient P=100.8kPa, Tair=26.9C, Tdew=-1.59C Zero Air at 800psi, AMP CA04396 408.97 at 550 Tdewspan=16.0 CO2 zero 0.9393 -> 0.9393 Offset: -0.1ppm CO2 span 1.016 -> 1.0171 Offset: -0.94ppm H2O zero 0.9096 -> 0.9072 Offset: 2.7ppm H2O span 1.0143 -> 1.0147 Offset: -0.01ppm Notes: Ambient air very dry
2020-05-21 75H-2668 LI-7500A Installed Gas Analyzer CO2/H2O Installed at Sherman Barn to replace sn 75H-2669 for calibration.
2020-07-09 75H-2668 LI-7500A Removed Gas Analyzer CO2/H2O Removed from Sherman Barn with its decommissioning and given to Sabina of Hydrofocus. The DWR grant ended at the end of June 2020, and Bryan requested we hand off a tower to Hydrofocus.
2018-10-18 75H-2669 LI-7500A Installed Gas Analyzer CO2/H2O Installed on Sherman Barn to replace sn 0041 for calibration
2019-04-15 75H-2669 LI-7500A Calibration Lab Last calibration: 2018-10-10 AGC=103.8 SDM=0 DAC1=- DAC2=- Ambient P=100.16kPa, Tair=24.85C, Tdew=6.62C Zero Air 369068 at 1150psi, AMP CA04396 408.97 at 800 Tdewspan=15.99 CO2 zero 0.9188 -> 0.9186 Offset: -1.10ppm CO2 span 1.0106 -> 1.0134 Offset: -1.52ppm H2O zero 0.8992 -> 0.8991 Offset: -0.3ppm H2O span 1.0034 -> 1.0017 Offset: 0.03ppm Notes: Very stable - Joe
2019-01-31 75H-2669 LI-7500A Removed Gas Analyzer CO2/H2O Removed from SB and brought back to the lab for calibration. It was replaced by 75H-2180.
2019-04-15 75H-2669 LI-7500A Calibration Lab Last calibration: 2018-10-10 AGC=103.8 SDM=0 DAC1=- DAC2=- Ambient P=100.16kPa, Tair=24.85C, Tdew=6.62C Zero Air 369068 at 1150psi, AMP CA04396 408.97 at 800 Tdewspan=15.99 CO2 zero 0.9188 -> 0.9186 Offset: -1.10ppm CO2 span 1.0106 -> 1.0134 Offset: -1.52ppm H2O zero 0.8992 -> 0.8991 Offset: -0.3ppm H2O span 1.0034 -> 1.0017 Offset: 0.03ppm Notes: Very stable - Joe
2019-04-18 75H-2669 LI-7500A Installed Gas Analyzer CO2/H2O Installed at SB to replace sn 75H-2180, which was brought back to the lab for calibration.
2019-08-01 75H-2669 LI-7500A Calibration Lab Last calibration: 2019-04-15 AGC=102.6 SDM= DAC1= DAC2= Ambient P=100.7kPa, Tair=22.7C, Tdew=14.1C Zero Air at 975psi, AMP CA4396 408.97 at 800 Tdewspan=20.00 CO2 zero 0.9186 -> .9178 Offset: -3.12ppm CO2 span 1.0134 -> 1.0125 Offset: .53ppm H2O zero 0.8991 -> .9019 Offset: 17.8ppm H2O span 1.0017 -> 1.015 Offset: -.3ppm Notes: When SS was 102.6, I reset it to 100. I uploaded the the calibration file from 2018-10-24 and copied in the manual coefficients for CO2 span, CO2 zero, H2O span, and H2O zero from the database's record of the most recent calibration on 2019-04-15.
2019-07-25 75H-2669 LI-7500A Removed Gas Analyzer CO2/H2O Removed from SB and brought back to the lab for calibration. it was replaced by sn 75H-2667.
2019-08-01 75H-2669 LI-7500A Calibration Lab Last calibration: 2019-04-15 AGC=102.6 SDM= DAC1= DAC2= Ambient P=100.7kPa, Tair=22.7C, Tdew=14.1C Zero Air at 975psi, AMP CA4396 408.97 at 800 Tdewspan=20.00 CO2 zero 0.9186 -> .9178 Offset: -3.12ppm CO2 span 1.0134 -> 1.0125 Offset: .53ppm H2O zero 0.8991 -> .9019 Offset: 17.8ppm H2O span 1.0017 -> 1.015 Offset: -.3ppm Notes: When SS was 102.6, I reset it to 100. I uploaded the the calibration file from 2018-10-24 and copied in the manual coefficients for CO2 span, CO2 zero, H2O span, and H2O zero from the database's record of the most recent calibration on 2019-04-15.
2020-02-06 75H-2669 LI-7500A Installed Gas Analyzer CO2/H2O Installed at Sherman Barn to replace sn 75H-2668, which was brought back to the lab for calibration.
2020-05-22 75H-2669 LI-7500A Calibration Lab Last calibration: 2020-02-04 AGC=100.7 SDM=0 DAC1=- DAC2=- Ambient P=100.2kPa, Tair=23.5C, Tdew=8.3C Zero Air at 750psi, AMP CA04396 408.97 at 600 Tdewspan=16.0 CO2 zero 0.9184 -> 0.9185 Offset: 0.44ppm CO2 span 1.016 -> 1.0124 Offset: 1.78ppm H2O zero 0.9006 -> 0.8999 Offset: -4.3ppm H2O span 1.0101 -> 1.0094 Offset: 0.01ppm Notes:
2020-05-21 75H-2669 LI-7500A Removed Gas Analyzer CO2/H2O Removed from Sherman Barn and brought back to the lab for calibration. Replaced by sn 75H-2668.
2020-05-22 75H-2669 LI-7500A Calibration Lab Last calibration: 2020-02-04 AGC=100.7 SDM=0 DAC1=- DAC2=- Ambient P=100.2kPa, Tair=23.5C, Tdew=8.3C Zero Air at 750psi, AMP CA04396 408.97 at 600 Tdewspan=16.0 CO2 zero 0.9184 -> 0.9185 Offset: 0.44ppm CO2 span 1.016 -> 1.0124 Offset: 1.78ppm H2O zero 0.9006 -> 0.8999 Offset: -4.3ppm H2O span 1.0101 -> 1.0094 Offset: 0.01ppm Notes:
2018-07-11 TG1-0357 LI-7700 Installed Gas Analyzer CH4 Installed at Sherman Barn on the scaffolding frame in the southwest corner of the cow fence. This is the initial set up of the site.
2019-04-11 TG1-0357 LI-7700 Calibration Gas Analyzer CH4 Calibration Location: Sherman Barn Last calibration: 2018-07-10 RSSI=70%, Optical RH=3% Ambient P=101.9kPa, Tair=16.5C Zero Air M4002 at 1800psi, AMP CA03923 3.121 at 1800 CH4 Zero Offset: 0.035ppm CH4 Span Offset: -0.002ppm Notes: Windy 3m/s to 7.5m/s
2019-09-04 TG1-0357 LI-7700 Upgrade Gas Analyzer CH4 We updated the 7700 firmware from v1.0.21 to v1.0.27. To do this we had to change the 7700’s IP address to 192.168.2.77 – it was still at the default 169 subnet
2019-10-31 TG1-0357 LI-7700 Calibration Gas Analyzer CH4 Calibration Location: Sherman Barn Last calibration: 2019-04-11 RSSI=68%, Optical RH=8% Ambient P=102.8kPa, Tair=12.1C Zero Air at 1000psi, AMP CA03923 3.121 at 1050 CH4 Zero Offset: 0.020ppm CH4 Span Offset: -0.011ppm Notes: Power was out earlier this week.
2020-05-21 TG1-0357 LI-7700 Calibration Gas Analyzer CH4 Calibration Location: Sherman Barn Last calibration: 2019-10-31 RSSI=80%, Optical RH=12% Ambient P=101.6kPa, Tair=19.2C Zero Air at 800psi, AMP CA3923 3.121 at 800 CH4 Zero Offset: 0.036ppm CH4 Span Offset: 0.036ppm Notes: Breezy day
2020-07-09 TG1-0357 LI-7700 Removed Gas Analyzer CH4 Removed from Sherman Barn with its decommissioning and given to Sabina of Hydrofocus. The DWR grant ended at the end of June 2020, and Bryan requested we hand off a tower to Hydrofocus.