Equipment and Calibration

Equipment history at Hill Slough

| ALL | | Anemometer || Battery Bank || Camera || CO2 Probe || Conductivity Sensor || Data Logger || DO Sensor || Fluorometer || Forced Diffusion || Gas Analyzer || Heat Flux Plate || Interface || Multiplexer || Narrow Band || Net Radiometer || PAR Sensor || pH ORP || Precipitation || Range Finder || RH & Temp || Soil Moisture Probe || Solar Controller || Solar Panel || Water Level || Wiper |

SELECT a.recno, a.eh_date, a.eh_loc1, a.eh_loc2, IF(a.eh_loc2=:loc, a.eh_locref2, a.eh_locref1) AS ref, a.eh_locref2, a.eh_title, a.eh_notes, a.eh_id, b.serialno, b.equip, b.modelno FROM equiphist a INNER JOIN equiplist b ON a.eh_id=b.recno WHERE (a.eh_loc1=:loc1 OR a.eh_loc2=:loc2) AND b.equip=:eq ORDER BY b.serialno, a.eh_date

Date Serial No. Model Action Ref Notes
Forced Diffusion
2021-03-25 FD20190028 eosFD Installed Forced Diffusion A Installed at Hill Slough on the ground to the southwest of the Tower platform.
2021-07-20 FD20190028 eosFD Maintenance Forced Diffusion A Performance check done. Referenced reset to 400ppm. Average Flux: -0.0, 0.03 STDEV Average ATM: 172.2ppm, 34.6C Average Soil: 171.6ppm, 34.5C
2021-10-26 FD20190028 eosFD Removed Forced Diffusion A Removed after being underwater when tidal flooding began. The pump was still running.