Equipment and Calibration
Equipment history at Bouldin Alfalfa| ALL | | Anemometer || Barometer || Camera || Data Logger || Flux Analyzer || Forced Diffusion || Gas Analyzer || Heat Flux Plate || Infrared Temperature || Interface || Multiplexer || Narrow Band || Net Radiometer || PAR Sensor || Precipitation || Range Finder || RH & Temp || Soil Moisture Probe || Solar Controller || Solar Panel || Weather Station |
SELECT a.recno, a.eh_date, a.eh_loc1, a.eh_loc2, IF(a.eh_loc2=:loc, a.eh_locref2, a.eh_locref1) AS ref, a.eh_locref2, a.eh_title, a.eh_notes, a.eh_id, b.serialno, b.equip, b.modelno FROM equiphist a INNER JOIN equiplist b ON a.eh_id=b.recno WHERE (a.eh_loc1=:loc1 OR a.eh_loc2=:loc2) AND b.equip=:eq ORDER BY b.serialno, a.eh_date
Date | Serial No. | Model | Action | Ref | Notes |
---|---|---|---|---|---|
Forced Diffusion | |||||
2018-07-12 | FD20160001 | eosFD | Installed | Forced Diffusion A | Installed at BA under the scaffolding. Sampling frequency is 5 min. |
2018-07-26 | FD20160001 | eosFD | Removed | Forced Diffusion A | Removed from BA because it was not turning on after power cycling and brought back to the lab for testing. I think the chamber is fine, just needs to be off for 1 minute before being turned back on. |
2018-08-07 | FD20160001 | eosFD | Installed | Forced Diffusion A | Installed under the BA scaffolding over bare soil. |
2020-04-17 | FD20160001 | eosFD | Notes | Forced Diffusion A | I tried to add a collar to this FD chamber, but the collar did not fit. I ended up reinstalling it in the same location (NE corner under the tower) over bare soil. |
2020-09-23 | FD20160001 | eosFD | Removed | Forced Diffusion A | Removed from BA and brought back to the lab for testing because of noisy data. It was replaced by sn FD20180048. |
2020-12-18 | FD20160001 | eosFD | Installed | Forced Diffusion A | Installed to replace the FD Chamber with the high temperature startup problems. |
2021-07-20 | FD20160001 | eosFD | Maintenance | Forced Diffusion A | Performance check done. Referenced reset to 400ppm. Average Flux: -0.1, 0.07 STDEV Average ATM: 402.1ppm, 48.5C Average Soil: 398.1ppm, 48.2C |
2023-04-11 | FD20160001 | eosFD | Removed | Forced Diffusion A | Removed from BA underneath the tower. The data has been super noisy since mid March. The cable is still on the tower. |
2023-05-09 | FD20180026 | eosFD | Moved | Forced Diffusion A | Moved from BC to BA to replace sn FD20160001 which broke. We should have soil CO2 flux data from Whendee's chamber measurements at BC. |
2024-10-09 | FD20180026 | eosFD | Removed | Forced Diffusion A | Removed from BA tower and brought back to the lab. The BA field is getting plowed. Joe may use this FD chamber at Tonzi/Vaira. |
2020-10-05 | FD20180047 | eosFD | Installed | Forced Diffusion A | Installed at Bouldin Alfalfa replacing sn FD20180048 that has low CO2 concentration readings. Also added a relay to restart this chamber at midnight to avoid high temperature start up problem. |
2020-12-18 | FD20180047 | eosFD | Removed | Forced Diffusion A | Removed from Bouldin Alfalfa. This one has high temperature start up issues and was running on a relay to restart it every night. I removed the relay too but did not update the CR1000 program. |
2020-09-23 | FD20180048 | eosFD | Moved | Forced Diffusion A | Moved from BC to BA to replace sn FD20160001, which was brought back to the lab for testing |
2020-10-05 | FD20180048 | eosFD | Removed | Forced Diffusion A | Removed from Bouldin Alfalfa to check on it's low CO2 concentration readings. Replaced by sn FD20180047 |
2020-04-17 | FD20190028 | eosFD | Installed | Forced Diffusion B | Installed with a collar at BA under the tower (SE corner) over bare soil. |
2020-09-23 | FD20190028 | eosFD | Removed | Forced Diffusion B | Removed from BA and brought back to the lab for testing because of noisy data |