Equipment history at Gilbert Tract| ALL | | Anemometer || Barometer || Battery Bank || Camera || CH4 Probe || CO2 Probe || Conductivity Sensor || Data Logger || DO Sensor || Flowmeter || Fluorometer || Gas Analyzer || Heat Flux Plate || Infrared Temperature || Interface || Multiplexer || Narrow Band || Net Radiometer || PAR Sensor || pH ORP || Precipitation || Range Finder || RH & Temp || Solar Controller || Solar Panel || Water Level || Weather Station || Wiper |
SELECT a.recno, a.eh_date, a.eh_loc1, a.eh_loc2, IF(a.eh_loc2=:loc, a.eh_locref2, a.eh_locref1) AS ref, a.eh_locref2, a.eh_title, a.eh_notes, a.eh_id, b.serialno, b.equip, b.modelno FROM equiphist a INNER JOIN equiplist b ON a.eh_id=b.recno WHERE (a.eh_loc1=:loc1 OR a.eh_loc2=:loc2) AND b.equip=:eq ORDER BY b.serialno, a.eh_date
Date | Serial No. | Model | Action | Ref | Notes |
---|---|---|---|---|---|
Fluorometer | |||||
2022-11-03 | 21K102591 | EXO fDOM | Calibration | fDOM (QSU and RFU), Last cal: 10/31/2022 4:14:35 PM EXO sn 22C100485 EXO/Sensor Firmware: 1.0.86/3.0.0 UCSC sensors sn 591 and sn 592 were really off even though calibrated recently. Will accept calibration and just run them at the GT channel for a week to see if they drift. ,Standard,PreCal,PostCal,RawValue,Temp Pt1,0.00 QSU,0.51 QSU,0.01 QSU,0.00 QSU,16.967 °C Pt2,316.50 QSU,219.51 QSU,316.49 QSU,0.00 QSU,10.855 °C ,Standard,PreCal,PostCal,RawValue,Temp Pt1,0.00 RFU,0.12 RFU,0.00 RFU,0.00 RFU,15.878 °C Pt2,105.50 RFU,74.86 RFU,105.50 RFU,0.00 RFU,11.257 °C | |
2023-01-25 | 21K102591 | EXO fDOM | Removed | none | USGS removed this EXO from the Gilbert Tract channel and stashed it on the bank. Joe retrieved it from the site at the end of the day and brought it back to the lab. |
2022-11-03 | 21K102592 | EXO fDOM | Calibration | fDOM (QSU and RFU), Last cal: 10/31/2022 4:14:35 PM EXO sn 22C100485 EXO/Sensor Firmware: 1.0.86/3.0.0 UCSC sensors sn 591 and sn 592 were really off even though calibrated recently. Will accept calibration and just run them at the GT channel for a week to see if they drift. ,Standard,PreCal,PostCal,RawValue,Temp Pt1,0.00 QSU,1.87 QSU,0.00 QSU,0.00 QSU,16.796 °C Pt2,316.50 QSU,218.07 QSU,316.49 QSU,0.00 QSU,10.735 °C ,Standard,PreCal,PostCal,RawValue,Temp Pt1,0.00 RFU,0.20 RFU,0.00 RFU,0.00 RFU,15.801 °C Pt2,105.50 RFU,71.69 RFU,105.50 RFU,0.00 RFU,11.214 °C | |
2023-01-25 | 21K102592 | EXO fDOM | Removed | none | USGS removed this EXO from the Gilbert Tract channel and stashed it on the bank. Joe retrieved it from the site at the end of the day and brought it back to the lab. |
2023-01-25 | 22C100485 | EXO Turbidity | Removed | none | USGS removed this EXO from the Gilbert Tract channel and stashed it on the bank. Joe retrieved it from the site at the end of the day and brought it back to the lab. |
2022-05-06 | 22C100780 | EXO fDOM | Installed | none | Mckenna from USGS installed this at the GT breach site next to the other USGS water sensors in the channel. They plan to service the sensors every 3 months. |
2022-11-03 | 22C100780 | EXO fDOM | Calibration | fDOM (QSU and RFU), Last cal: 10/24/2022 8:08:51 AM EXO sn 22C100485 EXO/Sensor Firmware: 1.0.86/3.0.0 UCSC sensors sn 591 and sn 592 were really off even though calibrated recently. Will accept calibration and just run them at the GT channel for a week to see if they drift. ,Standard,PreCal,PostCal,RawValue,Temp Pt1,0.00 QSU,-1.28 QSU,0.00 QSU,0.00 QSU,16.696 °C Pt2,316.50 QSU,315.87 QSU,316.54 QSU,0.00 QSU,10.444 °C ,Standard,PreCal,PostCal,RawValue,Temp Pt1,0.00 RFU,0.11 RFU,0.00 RFU,0.00 RFU,10.610 °C Pt2,105.50 RFU,92.32 RFU,105.51 RFU,0.00 RFU,11.275 °C | |
2024-07-23 | 22C100780 | EXO fDOM | Removed | Returned to us from USGS. Was deployed with their EXO in the Gilbert Tract channel. | |
2024-08-16 | 22C100780 | EXO fDOM | Installed | fDOM | Installed with EXO3s in Gilbert Tract channel, co-located with the ADCP and channel miniDOT |
2024-10-17 | 22C100780 | EXO fDOM | Removed | fDOM | Removed from GT channel and brought back to the lab for calibration |
2024-10-23 | 22C100780 | EXO fDOM | Installed | fDOM | Installed in the GT channel after calibration |
2024-08-16 | 24F102410 | EXO Turbidity | Installed | Turbidity | Installed with EXO3s in Gilbert Tract channel, co-located with the ADCP and channel miniDOT |
2024-10-17 | 24F102410 | EXO Turbidity | Removed | Turbidity | Removed from GT channel and brought back to the lab for calibration |
2024-10-23 | 24F102410 | EXO Turbidity | Installed | Turbidity | Installed in the GT channel after calibration |