Equipment history at Hill Slough| ALL | | Anemometer || Battery Bank || Camera || CO2 Probe || Conductivity Sensor || Data Logger || DO Sensor || Fluorometer || Forced Diffusion || Gas Analyzer || Heat Flux Plate || Interface || Multiplexer || Narrow Band || Net Radiometer || PAR Sensor || pH ORP || Precipitation || Range Finder || RH & Temp || Soil Moisture Probe || Solar Controller || Solar Panel || Water Level || Wiper |
SELECT a.recno, a.eh_date, a.eh_loc1, a.eh_loc2, IF(a.eh_loc2=:loc, a.eh_locref2, a.eh_locref1) AS ref, a.eh_locref2, a.eh_title, a.eh_notes, a.eh_id, b.serialno, b.equip, b.modelno FROM equiphist a INNER JOIN equiplist b ON a.eh_id=b.recno WHERE (a.eh_loc1=:loc1 OR a.eh_loc2=:loc2) AND b.equip=:eq ORDER BY b.serialno, a.eh_date
Date | Serial No. | Model | Action | Ref | Notes |
---|---|---|---|---|---|
DO Sensor | |||||
2021-11-04 | 931528 | miniDOT-usb | Installed | DO Sensor | Installed on a yellow chain near the SW corner of the platform. I think the sensor window is about 10cm above the ground |
2021-12-01 | 931528 | miniDOT-usb | Removed | DO Sensor | Removed from HS and brought back to the lab to swap its batteries. |
2021-12-02 | 931528 | miniDOT-usb | Installed | DO Sensor | I reinstalled the MiniDot that we removed yesterday due to dead batteries. |
2022-09-29 | 931528 | miniDOT-usb | Removed | DO Sensor | Removed from the HS channel and brought back to the lab for cleaning. There are calcareous deposits (some sort of encrusting worm) on and around the sensing foil. I soaked it in distilled white vinegar for 3 days. The deposits came off easily with a soft toothbrush. I rinsed the sensor well under DI water. |
2022-10-05 | 931528 | miniDOT-usb | Installed | DO Sensor | Installed after cleaning and measurement check in the lab. |
2024-09-26 | 931528 | miniDOT-usb | Adjusted | DO Sensor | Set clock in the field +1628 seconds (+27.1 minutes) to match internet time (used my cell phone hotspot to connect the laptop to the internet) |