Equipment history at West Pond| ALL | | Anemometer || Barometer || Battery Bank || Camera || CO2 Probe || Conductivity Sensor || Data Logger || DO Sensor || Gas Analyzer || Interface || Multiplexer || Narrow Band || Net Radiometer || PAR Sensor || RH & Temp || Solar Controller || Solar Panel || Water Level || Weather Station |
SELECT a.recno, a.eh_date, a.eh_loc1, a.eh_loc2, IF(a.eh_loc2=:loc, a.eh_locref2, a.eh_locref1) AS ref, a.eh_locref2, a.eh_title, a.eh_notes, a.eh_id, b.serialno, b.equip, b.modelno FROM equiphist a INNER JOIN equiplist b ON a.eh_id=b.recno WHERE (a.eh_loc1=:loc1 OR a.eh_loc2=:loc2) AND b.equip=:eq ORDER BY b.serialno, a.eh_date
Date | Serial No. | Model | Action | Ref | Notes |
---|---|---|---|---|---|
Conductivity Sensor | |||||
2012-04-17 | 6096 | CS547A | Installed | Conductivity Sensor | Installed on post northwest of scaffolding at west pilot pond. Not running because of forgotten power connector. |
2020-05-29 | 6096 | CS547A | Calibration | Conductivity Sensor | Calibration in the field with DI water (0 mS) and a YSI solution (10,000 uS). Temperature-corrected mS/cm readings from the data logger, where the temperature correction occurs in the data logger program. In DI water: 0.00mS/cm, 20.9C In 10,000 uS calibration soln: 10.3mS/cm, 23.0C Calibration: (Corrected mScm) = 0.9709 * (Sensor mScm) |
2024-02-12 | 6096 | CS547A | Removed | Conductivity Sensor | Removed from WP with interface sn 3586 and brought back to the lab. Nothing wrong with the data (probably), I just wanted an extra sensor in the lab to play around with TempCoef. |