| ALL | | Anemometer || Barometer || Battery Bank || Camera || CO2 Probe || Conductivity Sensor || Data Logger || DO Sensor || Gas Analyzer || Heat Flux Plate || Interface || Multiplexer || Narrow Band || Net Radiometer || PAR Sensor || Precipitation || RH & Temp || Solar Controller || Solar Panel || Water Level || Weather Station |

SELECT a.recno, a.eh_date, a.eh_loc1, a.eh_loc2, IF(a.eh_loc2=:loc, a.eh_locref2, a.eh_locref1) AS ref, a.eh_locref2, a.eh_title, a.eh_notes, a.eh_id, b.serialno, b.equip, b.modelno FROM equiphist a INNER JOIN equiplist b ON a.eh_id=b.recno WHERE (a.eh_loc1=:loc1 OR a.eh_loc2=:loc2) AND b.equip=:eq ORDER BY b.serialno, a.eh_date

Date Serial No. Model Action Ref Notes
CO2 Probe
2017-08-16 GP20140016 GP-001 Installed CO2 Probe Installed at SW. I waterproofed the cable-sensor connection with multiple layers of liquid electrical tape and silicone caulk.
2018-10-18 GP20140016 GP-001 Removed CO2 Probe Removed from Sherman Wetland due to railed readings not responding the atmospheric CO2.
2018-11-01 GP20140084 GP-001 Installed CO2 Probe Installed at SW to replace sn GP20140016, which was removed b/c it was maxing out.
2021-01-17 GP20140084 GP-001 Removed CO2 Probe Removed from SW and brought back to the lab b/c the sensor is bad and only outputting noise
2016-05-24 GP20150101 GP-001 Installed CO2 Probe Installed on the Sherman Wetland tower at the initial set up of the site. In the air unil the site is flooded.
2017-08-16 GP20150101 GP-001 Removed CO2 Probe Removed from SW because it did not have a waterproof sensor-cable connection.