| ALL | | Anemometer || Barometer || Battery Bank || Camera || CO2 Probe || Conductivity Sensor || Data Logger || DO Sensor || Gas Analyzer || Heat Flux Plate || Interface || Multiplexer || Narrow Band || Net Radiometer || PAR Sensor || Precipitation || RH & Temp || Solar Controller || Solar Panel || Spectrometer || Water Level |

SELECT a.recno, a.eh_date, a.eh_loc1, a.eh_loc2, IF(a.eh_loc2=:loc, a.eh_locref2, a.eh_locref1) AS ref, a.eh_locref2, a.eh_title, a.eh_notes, a.eh_id, b.serialno, b.equip, b.modelno FROM equiphist a INNER JOIN equiplist b ON a.eh_id=b.recno WHERE (a.eh_loc1=:loc1 OR a.eh_loc2=:loc2) AND b.equip=:eq ORDER BY b.serialno, a.eh_date

Date Serial No. Model Action Ref Notes
CO2 Probe
2014-06-26 GP20140011 GP-001 Installed CO2 probe Installed at the East End tower.
2015-06-25 GP20140011 GP-001 Removed CO2 probe Removed this 2% probe from East End. Replaced by 4% probe sn GP20140021.
2018-04-20 GP20140019 GP-001 Installed CO2 probe Installed at EE. The previous probe had been brought back to the lab at the last site visit b/c it was unresponsive.
2019-01-03 GP20140019 GP-001 Removed CO2 probe Removed and brought back to the lab for testing.
2015-06-25 GP20140021 GP-001 Installed CO2 probe Installed this 4% probe at East End due to replaced 2% probe sn GP20140011.
2015-12-08 GP20140021 GP-001 Removed CO2 probe This 4% CO2 probe was removed from East End. It was replaced by an 8% probe sn GP20150109
2019-01-18 GP20150101 GP-001 Installed CO2 probe Installed at EE to replace sn GP20140019, which was brought back to the lab for testing.
2019-07-10 GP20150101 GP-001 Removed CO2 probe Removed from EE and brought back to the lab for further testing (it was maxing out, even when out of the water).
2015-12-08 GP20150109 GP-001 Installed CO2 probe This 8% probe was installed at East End to replace a 4% probe sn GP20140021
2016-11-21 GP20150109 GP-001 Removed CO2 probe Removed from East End where it was pulling about 6A of power and browning out the entire site.
2019-09-04 GP20160096 GP-001 Installed CO2 probe The CO2 probe attached to the plastic chain to hang next to the miniDot.
2019-09-18 GP20160096 GP-001 Removed CO2 probe Removed from East Pond because it was railed high even in air.
2019-10-02 GP20160096 GP-001 Installed CO2 probe Installed at EE off the north side of the tower.
2020-10-05 GP20160096 GP-001 Removed CO2 probe Removed from EE because it still rails high in air and in water. Will send it back to the factory