Equipment and Calibration

Equipment history at East End

| ALL | | Anemometer || Barometer || Battery Bank || Camera || CO2 Probe || Conductivity Sensor || Data Logger || DO Sensor || Gas Analyzer || Heat Flux Plate || Interface || Multiplexer || Narrow Band || Net Radiometer || PAR Sensor || Precipitation || RH & Temp || Solar Controller || Solar Panel || Spectrometer || Water Level |

SELECT a.recno, a.eh_date, a.eh_loc1, a.eh_loc2, IF(a.eh_loc2=:loc, a.eh_locref2, a.eh_locref1) AS ref, a.eh_locref2, a.eh_title, a.eh_notes, a.eh_id, b.serialno, b.equip, b.modelno FROM equiphist a INNER JOIN equiplist b ON a.eh_id=b.recno WHERE (a.eh_loc1=:loc1 OR a.eh_loc2=:loc2) AND b.equip=:eq ORDER BY b.serialno, a.eh_date

Date Serial No. Model Action Ref Notes
Barometer
2013-11-25 H0160069 PTB110 Moved Barometer The DWR Corn / Construction tower was decommissioned and rechristened at its new permanent location at the Twitchell East End Wetland
2015-01-05 H0160069 PTB110 Removed Barometer Removed from the site due to the tower falling over in the wind on 2014-12-30. Likely undamaged.
2015-01-20 H0160069 PTB110 Installed Barometer Reinstalled after being removed when the scaffolding fell over.
2016-06-24 H0160069 PTB110 Removed Barometer Removed from East End because it was reading low and erratically.
2016-08-31 H0160069 PTB110 Installed Barometer Reinstalled on the East End tower after calibration at Vaisala.
2016-09-13 H0160069 PTB110 Removed Barometer Was reading low in the field again (around 98kPa) even though it was just returned from Campbell. Removed from EE tower for testing in the lab.
2016-09-29 H0160069 PTB110 Installed Barometer Tested okay in the lab. Reinstalled to try again.
Export to BADM