Equipment history at Sherman Island

| ALL | | Anemometer || Barometer || CO2 Probe || CO2 Transmitter || Data Logger || Gas Analyzer || Heat Flux Plate || Interface || LED sensor || Moisture Point || Multiplexer || Net Radiometer || PAR Sensor || Precipitation || Pump || RH & Temp || Scroll Pump || Soil Moisture Probe || Solar Controller || Water Level |

SELECT a.recno, a.eh_date, a.eh_loc1, a.eh_loc2, IF(a.eh_loc2=:loc, a.eh_locref2, a.eh_locref1) AS ref, a.eh_locref2, a.eh_title, a.eh_notes, a.eh_id, b.serialno, b.equip, b.modelno FROM equiphist a INNER JOIN equiplist b ON a.eh_id=b.recno WHERE (a.eh_loc1=:loc1 OR a.eh_loc2=:loc2) AND b.equip=:eq ORDER BY b.serialno, a.eh_date

Date Serial No. Model Action Ref Notes
Anemometer
2009-06-26 000019 WMP 1352 Installed Anemometer 2ndTower Installed the the secondary tower to replace sn 074417 which has a firmware issue that reduces the response rate of the SOS/T channel.
2010-04-06 000019 WMP 1352 Moved Anemometer 2ndTower Moved along with the secondary tower from the Sherman cow pasture to the Sherman Levee location.
2015-05-12 000020 WMP 1352 Removed Anemometer Removed with the decommissioning of the Sherman Cow Pasture site.
2012-09-27 000603 SATI-3K Installed Installed just to the south of the existing Gill WMP for an inter-comparison. The serial data was added to the CR1000 on the RS-232 port and the program was changed to add a third table to record the ATI data.
2012-10-18 000603 SATI-3K Installed Anemometer ATI Removed from the Sherman tower and installed on the West Pond. The Sherman system was returned to configuration before installing the ATI. At West Pond, the ATI was installed northeast of the Gill WMP just west (and upwind) of the LI7500. The data is recorded by a separate CR1000 in the box that Robin supplied with the ATI.