Site | Date | Author | Summary |
Hill Slough |
(355) 2022-12-21 09:30:00 |
Joe Verfaillie |
Quick stop to collect data. Fixed CTD that had its sdi-12 data hung up, cold and foggy
|
Hill Slough |
(334) 2022-11-30 09:50:00 |
Daphne Szutu |
Regular data collection/cleaning, swapped tower batteries, calibrated 7700, swapped 7500 for calibration
|
Hill Slough |
(321) 2022-11-17 09:35:00 |
Daphne Szutu |
Regular data collection/cleaning, biofilms(?) on water surface, HMP fan broken--fixed
|
Hill Slough |
(306) 2022-11-02 17:20:00 |
Daphne Szutu |
Regular data collection/cleaning, cold and windy, rearranged water sensors in channel
|
Hill Slough |
(291) 2022-10-18 09:55:00 |
Daphne Szutu |
Regular data collection/cleaning, Calm day, Rinsed off EXO, Need more work on mounting water quality sensors
|
Hill Slough |
(278) 2022-10-05 09:15:00 |
Joe Verfaillie |
Quick stop to reinstall miniDot, moved water level and pCO2 sensors, EXO interface serial number
|
Hill Slough |
(272) 2022-09-29 09:10:00 |
Daphne Szutu |
Regular data collection/cleaning, only a slight breeze, redeployed EXO after calibration and wired to CR23x, removed miniDOT for cleaning, picam power cable corroded--fixed, cleaned precip bucket
|
Hill Slough |
(259) 2022-09-16 07:45:00 |
Daphne Szutu |
Regular data collection/cleaning, removed EXO to bring back to the lab for upgrades/calibration, sipper water sample, went to Rush Ranch with Kat and Balt, first fall precip expected in a few days, record-breaking heat wave last week.
|
Hill Slough |
(243) 2022-08-31 14:30:00 |
Joe Verfaillie |
Low tide, standard fieldwork, power for water measurements looks okay, sipper water sample stored
|
Hill Slough |
(236) 2022-08-24 14:40:00 |
Daphne Szutu |
Moved eosGP and CTD sensor to channel, Installed CR23X and solar panel/li battery at channel, Changed EXO from 2min to 5min samples, high tide and windy
|
Hill Slough |
(227) 2022-08-15 15:40:00 |
Joe Verfaillie |
Hot, EXO batteries okay, Low tide, top solar panel cleaned, CTD water level un-buried.
|
Hill Slough |
(216) 2022-08-04 15:00:00 |
Daphne Szutu |
Regular data collection/cleaning, only a few days of data on EXO--changed batteries, cleaned sensors, and restarted deployment
|
Hill Slough |
(207) 2022-07-26 16:00:00 |
Ariane Arias-Ortiz |
Changed EXO2 deployment configuration to fix the sonde's autonomous wiping (not wiping) and removed the silver 9-pin serial connector from the sonic cable
|
Hill Slough |
(194) 2022-07-13 09:45:00 |
Daphne Szutu |
Regular data collection/cleaning, super low water level, installed EXO sonde in channel, moved miniDOT to channel, tested batteries--removed 2 bad ones, new picam code, installed sipper for soil pore water samples
|
Hill Slough |
(179) 2022-06-28 07:20:00 |
Carlos Wang |
Regular data collection/cleaning, losing power overnight
|
Hill Slough |
(159) 2022-06-08 18:15:00 |
Daphne Szutu |
Regular data collection/cleaning, calibrate 7700, skipped camera download--next time
|
Hill Slough |
(147) 2022-05-27 09:15:00 |
Joe Verfaillie |
Standard data collection, checked 7700 wash cycle - seems okay, low tide.
|
Hill Slough |
(131) 2022-05-11 15:20:00 |
Daphne Szutu |
Regular data collection/cleaning, some upland plants greening up
|
Hill Slough |
(110) 2022-04-20 10:00:00 |
Daphne Szutu |
Quick stop to check power--seem ok, swapped 7500 for calibration, slightly lowered CTD sensor
|
Hill Slough |
(103) 2022-04-13 17:15:00 |
Daphne Szutu |
Regular data collection/cleaning, windy, dfPAR numbers too high--fixed?, CTD sensor needs to be lowered next time, maybe intermittent power issues coming into datalogger box--needs more attention
|
Hill Slough |
(90) 2022-03-31 09:50:00 |
Daphne Szutu |
Regular data collection/cleaning, tested CTD conductivity--moved probe outside of well, adjusted heights of miniDOT and dissolved CO2 sensors, removed NR-LITE, removed Campbell water level sensor, new datalogger program
|
Hill Slough |
(74) 2022-03-15 09:30:00 |
Daphne Szutu |
Regular data collection/cleaning, rained overnight, many birds, low tide, changed eddy clock +1min
|
Hill Slough |
(60) 2022-03-01 17:10:00 |
Daphne Szutu |
Regular data collection/cleaning, golden hour and nice clouds
|
Hill Slough |
(48) 2022-02-17 09:50:00 |
Daphne Szutu |
Regular data collection/cleaning, windy, dfPAR noisy--power cycled CR10X
|
Hill Slough |
(34) 2022-02-03 09:05:00 |
Daphne Szutu |
Regular data collection/cleaning, Manuela visiting, removed soil moisture probes, 7700 optics RH lower today
|
Hill Slough |
(19) 2022-01-19 09:40:00 |
Joe Verfaillie |
Updated Pi-cam program, updated CR1000 program - conductivity still wonky, 7700 optics RH high, tried to fix by adding putty to TC module.
|
Hill Slough |
(6) 2022-01-06 09:35:00 |
Daphne Szutu |
Regular data collection/cleaning, foggy, changed 7700 optics desiccant again, raised miniDOT slightly
|