Equipment and Calibration History
View Equipment |
Date: |
2022-10-31 |
From: |
Lab - none |
To: |
Lab - none |
Title: |
Repair |
Notes: |
One of the wires was broken off right at the base of the sensor. I was able to solder on a lead and spliced a longer cable on and the sensor seemed to respond to light. From 2022-10-27 to 2022-10-31 I ran an inter-comparison out the lab window with sn 030382. The relationship was tight (r2=0.9999) and the slope was about 1.06. sn 030382 had been inter-compared on the Tonzi tower recently but in the reflected position so the range was not optimal. I think both sensors are functional and should be inter-compared with a good reference in the field to determine appropriate coefficients. |
Entered by: |
Joe Verfaillie |
|
|