Date: |
2020-11-25 |
From: |
Lab - none |
To: |
Lab - none |
Title: |
Maintenance |
Notes: |
Renamed from "Unknown" to "TG1-0224" and rebooted. The optics RH was 22%, so I swapped the desiccant and flushed the headspace with zero air. The new optics RH was 11%. I will let it run for a few days, logging data, to see if there are any issues before I calibrate. |
Entered by: |
Daphne Szutu |
|
|