Date: |
2020-05-29 |
From: |
East End - Conductivity Sensor |
To: |
East End - Conductivity Sensor |
Title: |
Calibration |
Notes: |
Calibration in the field with DI water (0 mS) and a YSI solution (10,000 uS). Temperature-corrected mS/cm readings from the data logger, where the temperature correction occurs in the data logger program.
In DI water: 0.00mS/cm, 22.4C
In 10,000 uS calibration soln: 11.3mS/cm, 23.0C
Calibration: (Corrected mScm) = 0.8850 * (Sensor mScm) |
Entered by: |
Joe Verfaillie |
|
|