Date: |
2018-11-06 |
From: |
Lab - none |
To: |
Lab - none |
Title: |
Tested |
Notes: |
Connected to a CR6 and recorded 10s data. The sensor was maxing out (reading ~5V) on both 0-2% and 0-6% ranges. The connection between sensor and cable seemed secure and clean. The sensor should go back to the factory. |
Entered by: |
Daphne Szutu |
|
|