Equipment and Calibration History
History for Vaisala GMP222 H1630066
Rec No. | Date | Loc From | Title | Loc To | Notes |
---|---|---|---|---|---|
1887 | 2015-06-09 | Twitchell Alfalfa | Removed | Lab none | Removed from forced diffusion soil cell with upgrade of soil CO2 system |
1869 | 2014-08-13 | Lab none | Installed | Twitchell Alfalfa | Installed near the radiometers at Alfalfa in the forced diffusion soil cell |
1818 | 2014-08-07 | Sherman Island | Removed | Lab none | Removed with the rest of the soil CO2 system from the Sherman site to be redeployed at the Alfalfa site on Twitchell |
1457 | 2014-05-27 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.746 + -27.77; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 2238ppm, and 3056ppm. Ta=20.19C, P=101.3kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Madeleine. |
2087 | 2014-04-10 | Sherman Island | Removed | Lab none | Removed from Sherman profile A at -2cm, replaced with sn K1340005 |
1353 | 2014-03-19 | Lab none | Installed | Sherman Island | This 0-1% probes was installed at Sherman in Set A at 2cm depth. |
1437 | 2014-01-27 | Lab none | Calibration | Lab none | Trueppm = Probeppm*1.753 + -29.94; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 3056ppm, and 10000ppm. Ta=22.7C, P=102.1kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Patty. |
958 | 2013-04-04 | Lab none | Maintenance | Lab none | |
957 | 2013-03-28 | Mayberry CO2 Probe B | Removed | Lab none | We removed the probe since it was reading 5000mV even we were put the sensor in the air. |
907 | 2013-01-17 | Lab none | Installed | Mayberry CO2 Probe B | Installed in a teflon sock in the water in the tule patch |
1419 | 2013-01-08 | Lab none | Calibration | Lab none | Trueppm = Probeppm*0.9828 + 15.78; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 421ppm, 3056ppm, and 10000ppm. Ta=22.3C, P=102.5kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Sara. |